Remote Commands for the I/Q Analyzer
R&S
®
FSWP
146User Manual 1177.5856.02 ─ 07
Suffix:
<n>
.
Window
Setting parameters:
<WindowType> Type of measurement window you want to replace another one
with.
See LAYout:ADD[:WINDow]? on page 140 for a list of availa-
ble window types.
Example:
LAY:WIND2:REPL MTAB
Replaces the result display in window 2 with a marker table.
Usage: Setting only
8.7
Analyzing Results
For a comprehensive description of the commands required to analyze results (trace
and marker functionality), please refer to the R&S FSWP User Manual.
CALCulate<n>:MARKer<m>:SEARch.............................................................................. 146
CALCulate<n>:SGRam:LAYout........................................................................................147
CALCulate<n>:SPECtrogram:LAYout............................................................................... 147
CALCulate<n>:SPECtrogram:TRACe...............................................................................147
CALCulate<n>:SGRam:TRACe....................................................................................... 147
CALCulate<n>:MARKer<m>:SEARch <MarkRealImag>
This command selects the trace type a marker search is performed on.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<MarkRealImag> REAL
Marker search functions are performed on the real trace of the
"I/Q" measurement.
IMAG
Marker search functions are performed on the imaginary trace of
the "I/Q" measurement.
MAGN
Marker search functions are performed on the magnitude of the I
and Q data.
*RST: REAL
Example:
CALC4:MARK:SEAR IMAG
Manual operation: See " Branch for Peaksearch " on page 70