Remote Commands for the I/Q Analyzer
R&S
®
FSWP
128User Manual 1177.5856.02 ─ 07
TRACe:IQ:EGATe:LENGth <GateLength>
This command defines the gate length for gated measurements with the I/Q analyzer.
Parameters:
<GateLength> <numeric value>
Max = (440 MS * sample rate/200MHz) -1
pretrigger samples defined by TRACe:IQ:SET;
sample rate defined by TRACe:IQ:SRATe)
Range: 1…Max (samples)
*RST: 100
Example:
TRAC:IQ:EGAT:LENG 2000
TRACe:IQ:EGATe:NOF <Number>
This command defines the number of gate periods after the trigger signal for gated
measurements with the I/Q analyzer.
Parameters:
<Number> Range: 1 to 1023
*RST: 1
Example:
TRAC:IQ:EGAT:NOF 2
TRACe:IQ:EGATe:TYPE <Type>
This command selects the gate mode for gated measurements with the I/Q analyzer.
Note: The IF power trigger holdoff time is ignored if you are using the "Level" gate
mode in combination with an IF Power trigger.
Parameters:
<Type> LEVel
EDGE
*RST: EDGE
Example:
TRAC:IQ:EGAT:TYPE LEV
8.6.11 Configuring Data Acquisition
Useful commands for I/Q data acquisition described elsewhere
●
[SENSe:]SWEep:COUNt on page 93
●
[SENSe:]SWEep[:WINDow<n>]:POINts on page 94
●
[SENSe<n>:]SWEep:TIME on page 94
[SENSe:]IQ:BWIDth:MODE.............................................................................................129
[SENSe:]IQ:BWIDth:RESolution...................................................................................... 129
[SENSe:]IQ:FFT:ALGorithm.............................................................................................130
[SENSe:]IQ:FFT:LENGth.................................................................................................130
[SENSe:]IQ:FFT:WINDow:LENGth...................................................................................131
Configuring I/Q Analyzer Measurements