Remote Commands for the I/Q Analyzer
R&S
®
FSWP
93User Manual 1177.5856.02 ─ 07
ON | 1
Switches the function on
Example:
TRAC:IQ ON
Switches on acquisition of I/Q data.
TRAC:IQ:AVER ON
Enables averaging of the I/Q measurement data.
TRAC:IQ:AVER:COUN 10
Selects averaging over 10 data sets.
TRAC:IQ:DATA?
Starts the measurement and reads out the averaged data.
[SENSe:]SWEep:COUNt <SweepCount>
This command defines the number of measurements that the application uses to aver-
age traces.
In continuous measurement mode, the application calculates the moving average over
the average count.
In single measurement mode, the application stops the measurement and calculates
the average after the average count has been reached.
Example:
SWE:COUN 64
Sets the number of measurements to 64.
INIT:CONT OFF
Switches to single measurement mode.
INIT;*WAI
Starts a measurement and waits for its end.
Manual operation: See " Sweep/Average Count " on page 64
[SENSe:]SWEep:COUNt:CURRent?
This query returns the current number of started sweeps or measurements. This com-
mand is only available if a sweep count value is defined and the instrument is in single
sweep mode.
Return values:
<CurrentCount>
Example:
SWE:COUNt 64
Sets sweep count to 64
INIT:CONT OFF
Switches to single sweep mode
INIT
Starts a sweep (without waiting for the sweep end!)
SWE:COUN:CURR?
Queries the number of started sweeps
Usage: Query only