A measurement is started immediately with the default settings when the channel is
activated.
Different remote modes available
In remote control, two different modes for the I/Q Analyzer measurements are availa-
ble:
●
A quick mode for pure data acquisition
This mode is activated by default with the TRACe:IQ[:STATe] command. The
evaluation functions are not available; however, performance is slightly improved.
●
A more sophisticated mode for acquisition and analysis.
This mode is activated when a new channel is opened for the I/Q Analyzer applica-
tion (INST:CRE:NEW/ INST:CRE:REPL) or by an additional command (see
TRACe:IQ:EVAL on page 86).
CALCulate<n>:IQ:MODE.................................................................................................. 82
INSTrument:CREate:DUPLicate........................................................................................ 83
INSTrument:CREate[:NEW].............................................................................................. 83
INSTrument:CREate:REPLace.......................................................................................... 83
INSTrument:DELete......................................................................................................... 84
INSTrument:LIST?........................................................................................................... 84
INSTrument:REName.......................................................................................................85
INSTrument[:SELect]........................................................................................................85
SYSTem:PRESet:CHANnel[:EXEC]................................................................................... 86
TRACe:IQ:EVAL.............................................................................................................. 86
TRACe:IQ[:STATe]........................................................................................................... 86
CALCulate<n>:IQ:MODE <EvalMode>
This command defines whether the captured I/Q data is evaluated directly, or if it is
converted (via FFT) to spectral or time data first.
It is currently only available for I/Q Analyzer slave applications in multistandard mode
(not the MSRA Master).
Suffix:
<n>
.
irrelevant
Parameters:
<EvalMode> TDOMain
Evaluation in time domain (zero span).
FDOMain
Evaluation in frequency domain.
IQ
Evaluation using I/Q data.
Activating I/Q Analyzer Measurements