Remote Commands for the I/Q Analyzer
R&S
®
FSWP
118User Manual 1177.5856.02 ─ 07
8.6.7 Configuring the Frequency
CALCulate<n>:MARKer<m>:FUNCtion:CENTer................................................................ 118
[SENSe:]FREQuency:CENTer......................................................................................... 118
[SENSe:]FREQuency:CENTer:STEP................................................................................118
[SENSe:]FREQuency:CENTer:STEP:AUTO...................................................................... 119
[SENSe:]FREQuency:OFFSet......................................................................................... 119
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
This command matches the center frequency to the frequency of a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Suffix:
<n>
.
Window
<m> Marker
Example:
CALC:MARK2:FUNC:CENT
Sets the center frequency to the frequency of marker 2.
[SENSe:]FREQuency:CENTer <Frequency>
This command defines the center frequency.
Parameters:
<Frequency> The allowed range and f
max
is specified in the data sheet.
*RST: fmax/2
Default unit: Hz
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Manual operation: See " Center Frequency " on page 53
[SENSe:]FREQuency:CENTer:STEP <StepSize>
This command defines the center frequency step size.
You can increase or decrease the center frequency quickly in fixed steps using the
SENS:FREQ UP AND SENS:FREQ DOWN commands, see [SENSe:]FREQuency:
CENTer on page 118.
Parameters:
<StepSize> f
max
is specified in the data sheet.
Range: 1 to fMAX
*RST: 0.1 x span
Default unit: Hz
Configuring I/Q Analyzer Measurements