EasyManua.ls Logo

Rohde & Schwarz FSWP Series - Configuring Data Inputs and Outputs

Rohde & Schwarz FSWP Series
167 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration
R&S
®
FSWP
41User Manual 1177.5856.02 ─ 07
It is not available in the Spectrum application, only in applications that process I/Q
data, such as the I/Q Analyzer or optional applications.
For details, see the description in the R&S FSWP I/Q Analyzer User Manual ("Import-
ing and Exporting I/Q Data").
Note: Storing large amounts of I/Q data (several Gigabytes) can exceed the available
(internal) storage space on the R&S FSWP. In this case, it can be necessary to use an
external storage medium.
Note: Secure user mode.
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256 MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S FSWP base unit user manual.
Remote command:
MMEMory:STORe<n>:IQ:STATe on page 148
MMEMory:STORe<n>:IQ:COMMent on page 148
5.3 Configuring Data Inputs and Outputs
The R&S FSWP can analyze signals from different input sources and provide various
types of output (such as video or trigger signals).
Inputs...................................................................................................................... 41
Outputs....................................................................................................................45
5.3.1 Inputs
The I/Q Analyzer supports several input sources.
RF Input.................................................................................................................. 41
Power Sensors........................................................................................................43
External Generators................................................................................................43
Probes.....................................................................................................................44
External Mixers....................................................................................................... 44
Settings for Input from I/Q Data Files......................................................................44
5.3.1.1 RF Input
Access: "Overview" > "Input/Frontend" > "Input Source" > "Radio Frequency"
The remote commands required to configure inputs are described in Chapter 8.6.3,
"Configuring Level Characteristics", on page 111, Chapter 8.6.4, "Configuring the
Configuring Data Inputs and Outputs

Table of Contents

Other manuals for Rohde & Schwarz FSWP Series