How to Work with I/Q Data
R&S
®
FSWP
73User Manual 1177.5856.02 ─ 07
● "Measurement Time" how long the data is to be captured
● "Record Length" : the number of samples to be captured (also defined by sam-
ple rate and measurement time)
8. Select the "Display Config" button and select up to six displays that are of interest
to you.
Arrange them on the display to suit your preferences.
9. Exit the SmartGrid mode.
10. Start a new sweep with the defined settings.
a)
Select the Sequencer icon (
) from the toolbar.
b) Set the Sequencer state to "Off" .
c) Select the [RUN SINGLE] key.
7.1.2 How to Analyze Data in the I/Q Analyzer
1. Select the [MODE] key and select the "I/Q Analyzer" application.
2. Select the "Overview" softkey to display the "Overview" for an I/Q Analyzer mea-
surement.
3. Select the "Display Config" button and select up to six displays that are of interest
to you.
Arrange them on the display to suit your preferences.
4. Exit the SmartGrid mode and select the "Overview" softkey to display the "Over-
view" again.
5. Select the "Analysis" button in the "Overview" to make use of the advanced analy-
sis functions in the displays.
● Configure a trace to display the average over a series of sweeps (on the
"Trace" tab; if necessary, increase the "Average Count" ).
● Configure markers and delta markers to determine deviations and offsets within
the signal (on the "Marker" tab).
7.2 How to Export and Import I/Q Data
I/Q data can only be exported in applications that process I/Q data, such as the I/Q
Analyzer or optional applications.
Capturing and exporting I/Q data
1. Press the [PRESET] key.
How to Export and Import I/Q Data