Remote Commands for the I/Q Analyzer
R&S
®
FSWP
89User Manual 1177.5856.02 ─ 07
INITiate<n>:CONMeas
This command restarts a (single) measurement that has been stopped (using ABORt)
or finished in single measurement mode.
The measurement is restarted at the beginning, not where the previous measurement
was stopped.
As opposed to INITiate<n>[:IMMediate], this command does not reset traces in
maxhold, minhold or average mode. Therefore it can be used to continue measure-
ments using maxhold or averaging functions.
Suffix:
<n>
.
irrelevant
Example:
INIT:CONT OFF
Switches to single measurement mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the measurement counter to 20 measurements.
INIT;*WAI
Starts the measurement and waits for the end of the 20 mea-
surements.
INIT:CONM;*WAI
Continues the measurement (next 20 measurements) and waits
for the end.
Result: Averaging is performed over 40 measurements.
Manual operation:
See " Continue Single Sweep " on page 65
INITiate<n>:CONTinuous <State>
This command controls the measurement mode for an individual channel.
Note that in single measurement mode, you can synchronize to the end of the mea-
surement with *OPC, *OPC? or *WAI. In continuous measurement mode, synchroniza-
tion to the end of the measurement is not possible. Thus, it is not recommended that
you use continuous measurement mode in remote control, as results like trace data or
markers are only valid after a single measurement end synchronization.
For details on synchronization see the "Remote Basics" chapter in the R&S FSWP
User Manual.
If the measurement mode is changed for a channel while the Sequencer is active the
mode is only considered the next time the measurement in that channel is activated by
the Sequencer.
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF | 0 | 1