Remote Commands for the I/Q Analyzer
R&S
®
FSWP
88User Manual 1177.5856.02 ─ 07
TRACe:IQ:AVERage:COUNt............................................................................................. 92
[SENSe:]AVERage<n>[:STATe<t>].....................................................................................92
TRACe:IQ:AVERage[:STATe]............................................................................................ 92
[SENSe:]SWEep:COUNt...................................................................................................93
[SENSe:]SWEep:COUNt:CURRent?..................................................................................93
[SENSe:]SWEep[:WINDow<n>]:POINts..............................................................................94
[SENSe<n>:]SWEep:TIME................................................................................................94
SYSTem:SEQuencer........................................................................................................94
ABORt
This command aborts the measurement in the current channel and resets the trigger
system.
To prevent overlapping execution of the subsequent command before the measure-
ment has been aborted successfully, use the *OPC? or *WAI command after ABOR and
before the next command.
For details see the "Remote Basics" chapter in the R&S FSWP User Manual.
To abort a sequence of measurements by the Sequencer, use the INITiate:
SEQuencer:ABORt command.
Note on blocked remote control programs:
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish and the remote
channel to the R&S FSWP is blocked for further commands. In this case, you must
interrupt processing on the remote channel first in order to abort the measurement.
To do so, send a "Device Clear" command from the control instrument to the
R&S FSWP on a parallel channel to clear all currently active remote channels.
Depending on the used interface and protocol, send the following commands:
Visa: viClear()
GPIB: ibclr()
RSIB: RSDLLibclr()
Now you can send the ABORt command on the remote channel performing the mea-
surement.
Example:
ABOR;:INIT:IMM
Aborts the current measurement and immediately starts a new
one.
Example:
ABOR;*WAI
INIT:IMM
Aborts the current measurement and starts a new one once
abortion has been completed.
Usage: Event
Performing Measurements