Configuration
R&S
®
FSWP
64User Manual 1177.5856.02 ─ 07
Note: As opposed to previous versions of the I/Q Analyzer, the sweep settings are now
window-specific.
For some result displays, the sweep points may not be editable as they are determined
automatically, or restrictions may apply. For the I/Q vector result display, the number of
I/Q samples to record ( "Record Length" ) must be identical to the number of trace
points to be displayed ("Sweep Points"). Thus, the sweep points are not editable for
this result display. If the "Record Length" is edited, the sweep points are adapted auto-
matically. For record lengths outside the valid range of sweep points, i.e. less than 101
points or more than 100001 points, the diagram does not show valid results.
Using fewer than 4096 sweep points with a detector other than "Auto Peak" may lead
to wrong level results. For details see "Combining results - trace detector" on page 30.
Remote command:
[SENSe:]SWEep[:WINDow<n>]:POINts on page 94
Sweep/Average Count
Defines the number of measurements to be performed in the single sweep mode. Val-
ues from 0 to 200000 are allowed. If the values 0 or 1 are set, one measurement is
performed.
The sweep count is applied to all the traces in all diagrams.
If the trace modes "Average" , "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if "Sweep Count" = 0 (default), averaging is performed
over 10 measurements. For "Sweep Count" =1, no averaging, maxhold or minhold
operations are performed.
Remote command:
[SENSe:]SWEep:COUNt on page 93
TRACe:IQ:AVERage:COUNt on page 92
Continuous Sweep / Run Cont
After triggering, starts the sweep and repeats it continuously until stopped. This is the
default setting.
While the measurement is running, the "Continuous Sweep" softkey and the [RUN
CONT] key are highlighted. The running measurement can be aborted by selecting the
highlighted softkey or key again. The results are not deleted until a new measurement
is started.
Note: Sequencer. If the Sequencer is active, the "Continuous Sweep" softkey only con-
trols the sweep mode for the currently selected channel. However, the sweep mode
only takes effect the next time the Sequencer activates that channel, and only for a
channel-defined sequence. In this case, a channel in continuous sweep mode is swept
repeatedly.
Furthermore, the [RUN CONT] key controls the Sequencer, not individual sweeps.
[RUN CONT] starts the Sequencer in continuous mode.
Remote command:
INITiate<n>:CONTinuous on page 89
Data Acquisition and Bandwidth Settings