Remote Commands for the I/Q Analyzer
R&S
®
FSWP
91User Manual 1177.5856.02 ─ 07
Before this command can be executed, the Sequencer must be activated (see
SYSTem:SEQuencer on page 94).
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
INITiate:SEQuencer:MODE <Mode>
This command selects the way the R&S FSWP application performs measurements
sequentially.
Before this command can be executed, the Sequencer must be activated (see
SYSTem:SEQuencer on page 94).
A detailed programming example is provided in the "Operating Modes" chapter in the
R&S FSWP User Manual.
Note: In order to synchronize to the end of a sequential measurement using *OPC,
*OPC? or *WAI you must use SINGle Sequence mode.
For details on synchronization see the "Remote Basics" chapter in the R&S FSWP
User Manual.
Parameters:
<Mode> SINGle
Each measurement is performed once (regardless of the chan-
nel's sweep mode), considering each channels' sweep count,
until all measurements in all active channels have been per-
formed.
CONTinuous
The measurements in each active channel are performed one
after the other, repeatedly (regardless of the channel's sweep
mode), in the same order, until the Sequencer is stopped.
CDEFined
First, a single sequence is performed. Then, only those channels
in continuous sweep mode (INIT:CONT ON) are repeated.
*RST:
CONTinuous
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
Performing Measurements