Remote Commands for the I/Q Analyzer
R&S
®
FSWP
123User Manual 1177.5856.02 ─ 07
TRIGger[:SEQuence]:SLOPe <Type>
This command selects the trigger slope.
Parameters:
<Type> POSitive | NEGative
POSitive
Triggers when the signal rises to the trigger level (rising edge).
NEGative
Triggers when the signal drops to the trigger level (falling edge).
*RST: POSitive
Example:
TRIG:SLOP NEG
Manual operation: See " Slope " on page 57
TRIGger[:SEQuence]:SOURce <Source>
This command selects the trigger source.
For details on trigger sources see " Trigger Source " on page 55.
Note on external triggers:
If a measurement is configured to wait for an external trigger signal in a remote control
program, remote control is blocked until the trigger is received and the program can
continue. Make sure this situation is avoided in your remote control programs.
Parameters:
<Source> IMMediate
Free Run
EXT | EXT2
Trigger signal from one of the "Trigger Input/Output" connectors.
Note: Connector must be configured for "Input".
RFPower
First intermediate frequency
(Frequency and time domain measurements only.)
IFPower
Second intermediate frequency
IQPower
Magnitude of sampled I/Q data
For applications that process I/Q data, such as the I/Q Analyzer
or optional applications.
VIDeo
Video mode is available in the time domain and only in the
Spectrum application.
BBPower
Baseband power (for digital input via the optional Digital Base-
band Interface
PSEN
External power sensor
Configuring I/Q Analyzer Measurements