Configuration
R&S
®
FSWP
49User Manual 1177.5856.02 ─ 07
Shifting the Display ( Offset ) ← Reference Level
Defines an arithmetic level offset. This offset is added to the measured level. In some
result displays, the scaling of the y-axis is changed accordingly.
Define an offset if the signal is attenuated or amplified before it is fed into the
R&S FSWP so the application shows correct power results. All displayed power level
results are shifted by this value.
The setting range is ±200 dB in 0.01 dB steps.
Note, however, that the internal reference level (used to adjust the hardware settings to
the expected signal) ignores any "Reference Level Offset" . Thus, it is important to
keep in mind the actual power level the R&S FSWP must handle. Do not rely on the
displayed reference level (internal reference level = displayed reference level - offset).
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet on page 112
Unit ← Reference Level
The R&S FSWP measures the signal voltage at the RF input.
In the default state, the level is displayed at a power level of 1 mW (= dBm). Via the
known input impedance (50 Ω or 75 Ω, see " Impedance " on page 42), conversion to
other units is possible.
The following units are available and directly convertible:
dBm
dBmV
dBμV
dBμA
dBpW
Volt
Ampere
Watt
Remote command:
INPut<ip>:IMPedance on page 108
CALCulate<n>:UNIT:POWer on page 112
Setting the Reference Level Automatically ( Auto Level ) ← Reference Level
Automatically determines a reference level which ensures that no overload occurs at
the R&S FSWP for the current input data. At the same time, the internal attenuators
are adjusted so the signal-to-noise ratio is optimized, while signal compression and
clipping are minimized.
To determine the required reference level, a level measurement is performed on the
R&S FSWP.
If necessary, you can optimize the reference level further. Decrease the attenuation
level manually to the lowest possible value before an overload occurs, then decrease
the reference level in the same way.
You can change the measurement time for the level measurement if necessary (see "
Changing the Automatic Measurement Time ( Meastime Manual )" on page 67).
Remote command:
[SENSe:]ADJust:LEVel on page 139
Configuring the Amplitude