R&S ZVL GUI Reference
Trace Menu
Operating Manual 1303.6580.32-06
169
search ranges 1 to 10 is selected.
Search range properties
In contrast to the marker properties defined in the Marker and Search menus, the ten search ranges are
valid for the entire setup. This means that, once defined, each of them can be used for any trace in the
setup, irrespective of the channel that the trace belongs to.
The default search range of each new marker is Full Span. The analyzer provides greatest flexibility in
defining search ranges. In particular, two search ranges may overlap or even be identical. The search is
confined to the part of the search range that belongs to the sweep range.
The following example shows how search ranges can be used to search a trace for several bandpass
regions.
Remote control:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER
CALC
ulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STARt
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STOP
Search Result Off
Hides the info field with the results of a bandpass or a bandstop search and disables bandfiter Tracking.
The info field is displayed again (and tracking re-enabled) when a new bandfilter search is performed.
Bandfilter parameters
The info field contains the following search results:
Bandwidth is the n-dB bandwidth of the bandpass/bandstop region, where n is the selected x dB
Bandwidth. The bandwidth is equal to the difference between the Upper Bandwidth Edge (UBE)
and the Lower Bandwidth Edge (LBE).
Center is calculated as the geometric mean value of the LBE and UBE positions: f
Center
= sqrt (f
LBE
* f
UBE
).
Lower Band Edge is the closest frequency below the center frequency where the trace is equal to
the center value minus n dB.
Upper Band Edge is the closest frequency above the center frequency where the trace is equal to
the center value minus n dB.
The Quality Factor is the ratio between the Center frequency and the 3-dB Bandwidth; it does not
depend on the selected x dB Bandwidth. The Quality Factor is not available for bandstop