R&S ZVL GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
252
The Sweep menu provides the following settings:
• Sweep Type defines the position of sweep points (frequency values) in the sweep range.
• Number of Points sets the total number of measurement points over the sweep range.
• Frequency Step Size sets the distance between two consecutive frequency sweep points.
• Meas Delay delays the start of each sweep.
• Trigger selects the trigger mode for the measurement.
• Restart aborts the current measurement and restarts a new sweep sequence.
• If Single (All Chans) is selected the measurement is terminated after a single sweep or a group of
single sweeps defined in the Define Restart dialog.
• Define Restart opens a dialog to specify which channels are affected and how many sweeps are
repeated.
• Average On activates or de-activates the sweep average. With average on the measurement
results are averaged over a selected number of consecutive sweeps (Average Factor).
• Average Factor defines the number of consecutive sweeps to be averaged.
• Restart Average starts a new average cycle, clearing all previous results and thus eliminating their
effect on the new cycle. The new cycle is started as fast as possible; an ongoing sweep is
terminated immediately.
Sweep Type
The Sweep Type submenu defines the frequency sweep type and the position of the sweep points across
the sweep range.
• Lin Frequency is the default sweep type. The stimulus frequency is swept in equidistant steps