EasyManua.ls Logo

Rohde & Schwarz ZVL series - GPIB Interface (Option R&S FSL-B10); DC Power Supply (Option R&S FSL-B30)

Rohde & Schwarz ZVL series
624 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S ZVL Interfaces and Connectors
Rear Panel Connectors
Operating Manual 1303.6580.32-05 587
GPIB Interface (Option R&S FSL–B10)
GPIB interface in compliance with IEEE488 and SCPI. A computer for remote control can be connected
via this interface. To set up the connection, a shielded cable is recommended. See GPIB Bus Interface
above in this chapter.
DC Power Supply (Option R&S FSL-B30)
A DC power supply can be connected alternatively to the AC power supply; see DC Power Supply and
Battery in chapter 1. The connector is supplied with the other accessories of the option and is connected
according to the following diagram:
Pin No. Input (I) or
Output (O) or
Bidirectional (B)
Voltage Range Function
1 I
+11 V to +28 V
7 A to 2.7 A
P
ositive voltage
2 Ground
3 Not used
Caution!
DC power supply
The power supply (SELV) that is used must fulfill the requirements for
reinforced/double insulation for main supply circuits in accordance to DIN/EN/IEC
61010 (UL 61010B–1, CSA C22.2 No. 1010.1) or DIN/EN/IEC 60950 (UL 1950, CSA
C22.2 No. 950). It is recommended to fuse the DC power supply according to the
table below. Before switching on the instrument check the connection for correct
polarity.
In continuous operation the actual breaking current can differ from the rated breaking current. For fuse
selection take the characteristics of the fuse into account.

Table of Contents

Related product manuals