R&S ZVL GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
216
Remote
control:
[SENSe<Ch>:]FREQuency:STARt
[SENSe<Ch>:]FREQuency:STOP
[SENSe<Ch>:]FREQuency:CENTer
[SENSe<Ch>:]FREQuency:SPAN
Pwr Bw
The Pwr Bw submenu defines the power of the internal signal source and sets the step attenuators and
the IF bandwidths.
Power defines the power of the internal signal source.
Step Atten b1/b2...sets the attenuation for the received wave b1 or b2 respectively.
RF Off switches the internal and external power sources on (if checked) or off.
Meas Bandwidth selects the bandwidth of the IF measurement filter.
Power
Opens the numeric entry bar to set the power of the internal signal source (channel power).
The channel power determines the output power at the test ports; see info below.
The selected channel power applies to all source ports used in the active channel.
Remote control:
SOURce<Ch>:POWer[:LEVel][:IMMediate][:AMPlitude]