R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
383
CALCulate<Chn>:MARKer<Mk>:MINimum
Selects a search mode for marker no. <Mk> and initiates a minimum search. The marker must be created
before using CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
Note: This command is the ZVR-compatible equivalent of
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MAXimum.
<Chn>
Channel number used to identify the active trace.
<Mk>
Ma
r
ker number in the range 1 to 10.
Range [def. unit] – [–]
*RST value
SCPI, Command Types
Device-specific, no query.
CALCulate<Chn>:MARKer<Mk>:MODE CONTinuous | DISCrete
Sets marker no. <Mk> to continuous or discrete mode. The marker doesn't have to be created before
(CALCulate<Chn>:MARKer<Mk>[:STATe] ON), the mode can be assigned in advance.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
Parameters
CONTinuous – marker can be positioned on any point of the trace, and its
response values are obtained by interpolation.
DISCrete – marker can be set to discrete sweep points only.
*RST value CONT
SCPI, Command
Types
Device-specific, command or query.
Example:
Suppose that the active setup contains an active trace no. 1.
CALC:MARK:MODE DISC; MARK2 CONT
Create marker 1 in discrete mode and marker 2 in continuous mode.
CALC:MARK ON; MARK2 ON
Display the two markers. Due to the different modes the horizontal position can
be different.
CALCulate<Chn>:MARKer<Mk>:REFerence:MODE CONTinuous | DISCrete
Sets the reference marker to continuous or discrete mode. The marker doesn't have to be created before
(CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON), the mode can be assigned in advance.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored and may be
set to any value.
Parameters
CONTinuous – marker can be positioned on any point of the trace, and its
response values are obtained by interpolation.
DISCrete – marker can be set to discrete sweep points only.