R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
407
<Chn>
Channel number used to identify the active trace. If unspecified the numeric
suffix is set to 1.
S | Y | P
S-parameters, Y-parameters, Z-parameters
*RST value – (The initial representation of the trace is determined by means of
CALCulate<Ch>:PARameter:SDEFine)
SCPI, Command
Types
Device-specific, command or query.
Example:
*RST; CALC:PAR:MEAS 'Trc1'", '"Y-S22'
Select the converted admittance Y<–S22 as measurement parameter of the
default trace.
CALC:TRAN:COMP S
Convert the converted Y-parameter into an S-parameter.
CALCulate<Chn>:TRANsform:TIME:CENTer <numeric_value>
Defines the center time of the diagram in time domain.
<Chn>
Channel number used to identify the active trace
<numeric_value>
Center time of the diagram in time domain
Range [def. unit] -99.99999999999 s to +99.99999999999 s. The increment (parameters UP
or DOWN) is 0.1 ns. [s]
*RST value 1.5E-009 s
SCPI, Command
Types
Confirmed, command or query
Example:
*RST; CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable the time
domain transformation for the default trace.
CALC:TRAN:TIME:CENT 0; SPAN 5ns
Set the center time to 0 ns and the time span to 5 ns.
Note: If the x-axis is scaled in distance units (CALCulate<Chn>:TRANsform:TIME:XAXis
DISTance), then the center value is entered in m or foot (ft), depending on the Distance Unit
selected in the System Configuration dialog. The range and default value changes accordingly.
CALCulate<Chn>:TRANsform:TIME:DCHebyshev <numeric_value>
Sets the sideband suppression for the Dolph-Chebyshev window. The command is only available if a
Dolph-Chebyshev window is active (CALCulate<Chn>:TRANsform:TIME:WINDow DCHebyshev).