R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
416
CALC:TRAN:DTF:PEAK:STAT ON
Enable the fault limit check for the active trace.
CALC:TRAN:DTF:PEAK:COUN?
Query the number of detected faults. The network analyzer responds with an
integer number, e.g. 2.
CALC:TRAN:DTF:PEAK:DATA2?
Query the distance and peak response values of the second detected fault.
Possible response: 0.09615384615,-18.07471664 (for a distance of
0.09615384615 units of length and a peak response value of -18.07471664 dB).
CALCulate<Chn>:TRANsform:DTFault:POINts
Starts the automatic calculation of the number of points.
<Chn>
Channel number used to identify the active trace
Parameters
N/a
*RST value
SCPI, Command Types
Device-specific, no query
Example:
See CALCulate<Chn>:TRANsform:DTFault:STATe
CALCulate<Ch>:TRANsform:DTFault:SELect '<cable name>'
Selects a standard cable or a previously defined cable identified by its name.
<Ch>
Channel number. Note that the cable type is channel-specific (rather than
trace-specific).
'<cable name>'
Cable name used to identify the cable
*RST value
SCPI, Command
Types
Device-specific, command or query
Example:
See CALCulate<Chn>:TRANsform:DTFault:DEFine
CALCulate<Chn>:TRANsform:DTFault:SPAN <numeric_value>
Defines the span of the distance-to-fault diagram.
<Chn>
Channel number used to identify the active trace
<numeric_value>
Distance span
Range [def. unit] 1 mm to +15 m. The increment (parameters UP or DOWN) is 1 mm. [m or foot
(ft), depending on the Distance Unit selected in the System Configuration
dialog]
*RST value 10 m
SCPI, Command
Device-specific, command or query