R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
466
*RST value
SCPI,
Command
Types
Device-specific, no query.
Example:
Assume that the current setup contains two channels numbered 1
and 2, respectively, and that sweep segments have been defined
for channel no. 1.
:MMEM:STOR:SEGM 1,
'C:\R_S\Instr\user\Nwa\SweepSegments\Seg_Ch1.seglist'
Store the sweep segment definition of channel 1 to a sweep
segment file.
MMEM:LOAD:SEGM 2,
'C:\R_S\Instr\user\Nwa\SweepSegments\Seg_Ch1.seglist'
Load the previously created sweep segment file and use the
sweep segments for channel 2.
MMEMory:LOAD:STATe <numeric_value>,'<file_name>'
Loads configuration data from a specified device settings file and sets the analyzer to the corresponding
instrument state.
<numeric_value>
1 (the <numeric_value> is used for compatibility with the SCPI standard but is
ignored).
<file_name>
String parameter to specify the name and directory of the settings file to be
loaded. The default extension for settings files is *.zvl.dfl (see
MMEMory:STORe:STATe). This extension may be omitted when a settings file
is loaded.
If no path is specified the analyzer searches the current directory, to be queried
with MMEMory:CDIRectory?.
*RST value
SCPI, Command
Types
Confirmed, no query.
Example:
MMEM:STOR:STAT 1,'C:\R_S\Instr\user\Setup_0413.zvl.dfl'
Store the current configuration to the file
Setup_0413.zvl.dfl in the default directory for settings
files.
MMEM:STOR:STAT:NEXT
Store the current configuration to the file
Setup_0414.zvl.dfl in the default directory for settings
files.
MMEM:LOAD:STAT 1,'C:\R_S\Instr\user\Setup_0413.zvl.dfl'
Load the settings stored in Setup_0413.zvl.dfl.