R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
468
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<Ph_pt> '<file_name>'
Loads data from a specified two-port (*.sp2) Touchstone file defining the characteristics of the current
deembedded single ended port circuit model. The single ended port circuit model involving file import
(FIMPort) must be selected before using the command.
<Ch>
Channel number.
<Ph_pt>
Physical port number
'<file_name>'
String parameter to specify the name and directory of the loaded Touchstone file. If
no path is specified the analyzer searches the current directory, to be queried with
MMEMory:CDIRectory?.
*RST value –
SCPI,
Command
Types
Device-specific, no query.
Example:
CALC:TRAN:VNET:SEND:DEEM:TND FIMP
Select the Serial .s2p data, shunt L circuit model.
MMEM:LOAD:VNET:SEND:DEEM2
'C:\Rohde&Schwarz\NWA\VNETWORKS\Test.s2p'
Load a Touchstone file and assign it to the physical port no.
2.
MMEMory:LOAD:VNETworks<Ch>:SENDed:EMBedding<Ph_pt> '<file_name>'
Loads data from a specified two-port (*.sp2) Touchstone file defining the characteristics of the current
embedded single ended port circuit model. The single ended port circuit model involving file import
(FIMPort) must be selected before using the command.
<Ch>
Channel number.
<Ph_pt>
'<file_name>'
String parameter to specify the name and directory of the loaded Touchstone file. If
no path is specified the analyzer searches the current directory, to be queried with
MMEMory:CDIRectory?.
*RST value –
SCPI,
Command
Types
Device-specific, no query.
Example:
CALC:TRAN:VNET:SEND:EMB:TND FIMP
Select the Serial .s2p data, shunt L circuit model.
MMEM:LOAD:VNET:SEND:EMB2
'C:\Rohde&Schwarz\NWA\VNETWORKS\Test.s2p'
Load a Touchstone file and assign it to the physical port no.
2.