[SENSe<Ch>:]AVERage:COUNt <numeric_value>
Defines the number of consecutive sweeps to be combined for the sweep average (Average Factor).
<Ch>
Channel number.
<numeric_value>
Average factor
Range [def. unit] 1 to 1000 [1]
*RST value 10
SCPI, Command Types
Confirmed, command or query
Example:
AVER:COUN 15
Set the average factor for channel 1 to 15.
AVER ON
Enable the sweep average.
[SENSe<Ch>:]AVERage[:STATe] <Boolean>
Enable or disable the sweep average.
<Ch>
Channel number.
<Boolean>
ON | OFF - Enables or disables the automatic calculation of the sweep average
over the specified number of sweeps ([SENSe<Ch>:]AVERage:COUNt).
*RST value ON
SCPI,
Command
Types
Confirmed, command or query
Example:
AVER:COUN 15
Set the average factor for channel 1 to 15.
AVER ON
Enable the sweep average over the defined number of sweeps.
[SENSe<Ch>:]BANDwidth...
This subsystem sets the bandwidth of the IF measurement filter (resolution bandwidth) and the filter type.
The forms BANDwidth and BWIDth are equivalent.
[SENSe<Ch>:]BANDwidth|BWIDth[:RESolution] <bandwidth>
Defines the resolution bandwidth of the analyzer (Meas Bandwidth).
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
<bandwidth>
Resolution bandwidth
Range [def. unit] 10 Hz to 500 kHz. UP and DOWN increment/decrement the bandwidth in 1-2-5