R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
498
maintained. CORR:COLL:CONN4? returns N75FEMALE.
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs ALL | SINGle
Qualifies whether the connector types at the analyzer ports (but not their gender) are equal or
independent.
In the present firmware version, a calibration must be performed using the same connector type at all
ports. Use [SENSe<Ch>:]CORRection:COLLect:CONNection<port_no>:PORTs SINGle if you want
to perform a measurement with independent port connectors.
<Ch>
Channel number of the calibrated channel.
Parameters
ALL: Equal (uniform) connector types. If the connector type at one port is
changed, the connector type at all other ports is changed accordingly.
SINGle: Independent (possibly non-uniform) connector types at the ports.
<port_no>
Port numbers of the analyzer. This parameter has no effect because the setting
affects all ports.
*RST value ALL
SCPI, Command
Types
Device-specific, command or query.
Example:
CORR:COLL:CONN:PORTS SING
Select independent connector types at the ports.
CORR:COLL:CONN1 N50MALE; CONN4 N75FEMALE; CONN2?
Select independent connector types at ports 1 and 4. The connector type at port
2 is not changed; the query returns N50FEMALE.
[SENSe<Ch>:]CORRection:COLLect:DELete ['<cal_name>']
Deletes system error correction data generated and stored previously.
<Ch>
Channel number of the calibrated channel.
'<cal_name>'
Name of the calibration (string parameter) defined together with calibration type
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine. If nothing is specified
the analyzer deletes the last system error correction stored by means of
[SENSe<Ch>:]CORRection:COLLect:SAVE.
*RST value
SCPI,
Command
Types
Device-specific, no query
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the measurement results of
this standard.