R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
526
Decrease the power to –20 dBm.
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol <Boolean>
Qualifies whether or not the Power of the internal signal source can be set independently for each sweep
segment.
<Ch>
Channel number.
<Seg>
Sweep segment number.
<Boolean>
ON: The power can be set independently for each sweep segment.
OFF: The power in all sweep segments is equal to the internal source power for
unsegmented sweeps set via
SOURce<Ch>:POWer[:LEVel][:IMMediate][:AMPlitude].
*RST value OFF. The parameter is automatically switched to ON when a segment level is
entered using [SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel].
SCPI,
Command
Types
Device-specific, command or query
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus –
10 dBm internal source power.
SEGM:POW -20
Decrease the power to –20 dBm.
SEGM:POW:CONT OFF
Couple the powers in all segments and reset the power in segment no. 1 to the initial
value.
[SENSe<Ch>:]SEGMent<Seg>[:STATe] <Boolean>
Activates or deactivates the sweep segment <Seg>. Sweep points belonging to inactive segments only
are not measured
<Ch>
Channel number.
<Seg>
Sweep segment number.
<Boolean>
ON | OFF - Activates or deactivates the measurement in sweep segment
<Seg>.
*RST value ON
SCPI, Command
Types
Device-specific, command or query
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings.
SEGM OFF