R&S ZVL Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
536
*RST value
SCPI, Command
Types
Confirmed, query only
Example:
STAT:QUES:LIM1?
Query the EVENt part of the QUEStionable:LIMit1 register to check whether an
event has occurred since the last reading.
STATus:QUEStionable:LIMit<1|2>:NTRansition
Sets the negative transition filter. Setting a bit causes a 1 to 0 transition in the corresponding bit of the
associated condition register to cause a 1 to be written in the associated bit of the corresponding event
register.
<NRf>
0 to 65535 (decimal representation)
*RST value
(see also Reset Values of the Status Reporting System)
SCPI, Command Types
Confirmed, command or query
Example:
STAT:QUES:LIM2:NTR 6
Set bits no. 1 and 2 of the QUEStionable:LIMit2:NTRansition register
STATus:QUEStionable:LIMit<1|2>:PTRansition
Sets the positive transition filter. Setting a bit causes a 0 to 1 transition in the corresponding bit of the
associated condition register to cause a 1 to be written in the associated bit of the corresponding event
register.
<NRf>
0 to 65535 (decimal representation)
*RST value
(see also Reset Values of the Status Reporting System)
SCPI, Command Types
Confirmed, command or query
Example:
STAT:QUES:LIM2:PTR 6
Set bits no. 1 and 2 of the QUEStionable:LIMit2:PTRansition register
STATus:QUEue[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is identical to that of
SYSTem:ERRor[:NEXT]?
The entry consists of an error number and a short description of the error. Positive error numbers are
instrument-dependent. Negative error numbers are reserved by the SCPI standard; see section Error
Messages.
*RST value
SCPI, Command
Confirmed, query only