ZVL Getting Started
Reflection Measurements
Operating manual 1303.6538.62 - 05 46
4. In the first dialog of the wizard, select the calibration kit (here: ZV-
Z21) and the test port connector (here: N 50
(f), corresponding to
a male calibration standard), and click Next.
If you have not yet imported the exact cal kit data of your calibration kit,
you can use the typical data as shown above. Typical data provide an
approximate description of a calibration kit mode. To import the actual
(accurate) data of your kit, press Import Kit... and s
elect the appropriate
cal kit file.
The next dialog of the calibration wizard shows that only a single
calibration standard needs to be measured.
5. Click the box Short (m)... to initiate the measurement of the
connected short standard.
The analyzer performs a calibration sweep and displays a message