4-5
1. Select Option
Press
2. Select Option
3. See
4. Press Twice
5. Select ON
6. Press
7. Select Mode
Press
8. Press Twice
To Select Between Single Sensor Mode Or Diff/Ratio
Mode.
1. Follow steps shown in Figure 4-4 to select the
Diff/Ratio mode.
Background Information. The equations used for the
Diff/Ratio functions are:
SENSOR 2 (Product Conductivity)
1. RATIO = ––––––––––––––––––––––––––––––
SENSOR 1 (Feed Conductivity)
2. DIFF = SENSOR 2 - SENSOR 1
FIGURE 4-4 Selecting Diff/Ratio Display Mode.
––
l
PROGRAM
SENSOR
PRIMARY DISPLAY
INSTRUMENT ADDRESS
EXIT
F4
MAIN MENU
ALARMS
OUTPUT
PROGRAM
PROGRAM DISPLAY
DIFF/RATIO NEEDED OFF
AVAILABLE SENSOR 1
SEN1 K.01 UNITS
SEN2 K.01
PROGRAM DISPLAY
DIFF/RATIO NEEDED ON
AVAILABLE MODE:
RATIO RATIO
DIFF
––
l
NEXT
F3
NEXT
F3
MODEL 2700 SECTION 4.0
CONFIGURATION
PROGRAM DISPLAY
DIFF/RATIO NEEDED ON
AVAILABLE MODE:
SENSOR PROCESS UNITS AVAILABLE
TYPE USED IN
Pure Water
µS/cm Mohm-cm µS/m Kohm-m ppm-TDS
µS/cm R* Mohm-cm R* mS/m Mohm-m
µS/cm Mohm-cm Kohm-m µS/cm
Water µS/cm R* Kohm-cm R* Mohm-m mS/m
K=0.01
Mohm-cm R*
µS/cm Mohm-cm Mohm-m µS/m ppm-TDS
NaCl
µS/cm R* Mohm-cm R* Kohm-m mS/m
CATION
µS/cm Mohm-cm Kohm-m µS/m ppm-TDS
mS/cm R* Mohm-cm R* Mohm-m mS/m
Water
µS/cm Kohm-cm µS/cm R* ohm-m ppm-TDS
K=1.0
mS/cm Kohm-cm R* mS/m Kohm-m ppt-TDS
NaCl
µS/cm Kohm-cm mS/m ohm-m ppm-TDS
mS/cm Kohm-cm R* µS/cm R* Kohms-m ppt-TDS
Water
mS/cm Ohm-cm mS/m %CONC
mS/cm R* Ohm-m S/m Kohm-cm R*
4 ELEC.
NaCl
µS/cm Ohm-cm S/m
(K=0.85) %CONC Ohm-m mS/m
HCl mS/cm mS/cm R* S/m S/m R* %CONC
NaOH mS/cm mS/cm R* S/m S/m R* %CONC
TABLE 4-2. Stored Units For Conductivity Type Sensors.
* R-not temperature compensated measurement (“Raw”)
1. Select Option
Press
2. Select Option
3. See
4. Select
Press
5. Press
Sensor with
6. Select Units (Parameter
to make Primary)
Press
7. Press Twice
FIGURE 4-5. Selecting Single Sensor Display Mode.
––
l
PROGRAM
SENSOR
PRIMARY DISPLAY
INSTRUMENT ADDRESS
MAIN MENU
ALARMS
OUTPUT
PROGRAM
PROGRAM PRI DISPLAY
DIFF/RATIO NEEDED OFF
AVAILABLE SENSOR 1
SEN1 K.01 UNITS
SEN2 K.01
PROGRAM PRI DISPLAY
DIFF/RATIO NEEDED OFF
AVAILABLE SENSOR 1
mS/cm UNITS
DEG C mS/cm
––
l
NEXT
F3
––
l
EXIT
F4
SENSOR 2 (Product Conductivity)
3.
%PASS =
–––––––––– x 100
SENSOR 1 (Feed Conductivity)
4. % REJECT = 100 - (% PASS)
Sensor 1 and 2 are the sensors addressed as #1 and
#2 during start-up. To change a sensor address see
Section 4.9
To Select Single Sensor Mode.
1. Follow the steps shown in Figure 4-5.