EasyManua.ls Logo

Rosemount 2700 - Troubleshooting Guide

Rosemount 2700
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-3
MODEL 2700 SECTION 6.0
TROUBLESHOOTING
TABLE 6-2 Quick Troubleshooting Guide
SYMPTOM ACTION
Wrong temperature reading
Perform a Single point calibration. (Section 3.0) If cannot calibrate, sensor is bad.
Suspected temperature • Adjust the analyzer temperature to that at the sensor.
compensation problem. Insure Calibration is not beyond ± 2 Deg C of raw value.
Display pixel missing,
Remove sensor wiring; toggle power
Display not operating correctly.
• Replace display board.
Analyzer locks up; won’t respond.
• Remove sensor wiring; toggle power.
• Check Keypad cable connection. CAUTION: Cable terminal failure can result when
removed more than 5 times.
Replace electronics module.
• Toggle power.
Analyzer not responding to • Verify and clean ribbon cable connection on CPU board.
key presses.
Key press gives wrong • Check Keypad cable connection. CAUTION: Cable terminal failure can result
selection. when removed more than 5 times.
• Replace keyboard front panel assy.
Wrong or no current output. • Verify: Output is not being overloaded; remove load:
(Typical voltage across unloaded output is 14-15 Vdc) Current load is not more than 600
ohms. There are no breaks in connections. Wiring is correct. Operating ranges are
set correctly Whether or not Auto Ranging function is operative
• Replace electronics module.
No backlight on display, or • Verify power wiring.
indicators. • Toggle power
Blows Fuses. • Wrong fuse. Replace with P/N 9100178
• Replace electronics module.
Drifting or unstable Readings.
Air bubbles in process affecting reading. Remount sensor to decrease exposure
(also see suspected Temp. to air bubbles.
Comp. problem) • Check programming of Liquid Type.
• Check programming of units of measure.
• Incorrect sensor cell constant for application.
• Verify process temperature is stable.
• Electrical or grounding interference.
• Verify sensor works in a standard solution outside of the process.
Verify nearby water pumps motors are properly grounded.
• Verify that sensor shield never touches an Earth Ground.
• Run sensor cable through grounded metal conduit to reduce RFI effects.
Inaccurate or erroneous • Wrong sensor for measurement range.
readings. • Verify that the cell constant is correct for the measurement range
(conductivity).
• Verify Liquid Type selected (See Figure 4-2) (conductivity).
• Place sensor in known solution and check reading. If incorrect, sensor needs to be
replaced.
pH reading won’t change in
• Clean the electrode, check for cracks in the bulb.Replace electrode if
different buffers cracked. Check impedance. If zero, replace glass electrode.
pH sensor has sluggish
• Check the electrode; if still sluggish soak in 10% HCl for 1 hr.
response. Replace electrode if not rejuvenated by the HCl soak.
Will calibrate in buffers but
• Process problem or possible ground loop. Make sure that the shield wires do
not in process. NOT touch grounded metal. Disconnect output and relay connections.
6.3 TROUBLESHOOTING GUIDE
Table 6-2 lists typical symptoms that may be
encountered and suggested actions.
These are not meant to be all inclusive, but to provide a
starting point for some of the more common problems
that may occur.

Table of Contents

Related product manuals