EasyManua.ls Logo

rotork Mk3 - Setting up with the Setting Tool

rotork Mk3
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Foundation Fieldbus CP Option Card Technical Manual
78 of 116 Publication PUB089-005-00_0618
8.1 Setting up with the Setting Tool
The actuator includes a communication port for setting the actuator performance, limit switches etc.
via the Setting Tool. This communication link can be used to set some, but not all, of the actuator
specific Transducer Block parameters. The parameters will be supplied set to their factory default
values. These should still be checked with a configuration tool to ensure optimum actuator and
network compatibility. Refer to the relevant actuator’s “Full Configuration Manual” to see the access
route through the actuator menu screens to reach the settings that affect the FF CP card.
The Transducer block parameters that may be set by using the Setting Tool are listed below. The
available range for the deadband setting when using the setting tool is less than the range found using
the configuration tools.
FF TB
Parameter
Description
Range
Default
Value
50
Limited Range Position Minimum
0 100%
0%
51
Limited Range Position Maximum
0 100%
100%
52
Deadband
0.0 9.9%
5.0%
53
Hysteresis
0.0 9.9%
2.0%
55
Motion Inhibit Time
0 255 sec
5 sec
59
Comms loss action
0 No Action
1 Open
3 Close
5 Stop
7 Position
Any other value = off
0
60
Comms lost position
0-100%
0
61
Comms Fault timer
0-255s
10
62
Auxiliary Mask (Function)
0x00-0xFF
0x0F
Note:
Parameters may only be altered by the setting tool if the actuator control knob is set to the
Local or Local Stop position.

Table of Contents

Related product manuals