-45-
Electrical Adjustment
⁄1 Wind sensor calibration
1. Enter the service mode.
2. Select Group “240” and No. “2”. Confirm the Data
value is “1”.
0: Wind sensor function Disable
1: Wind sensor function Enable
3. Select Group “240” and No. “0”.
4. To start the adjustment, change the Data value from
“0” to “1”. After the auto-calibration completed, "OK"
will appear on the screen.
IMPORTANT
Before taking this adjustment, you need to replace the
air filter with new one or scroll up the filter sheet and
make sure that the cabinet, filter unit or projection lens is
securely installed.