Operation 31
Programming Tests
Select the programs in the menu “Programming tests: Program selection”. You can
select existing programs by entering the line numbers or by selecting them directly
on the touch screen. The selected line is highlighted in blue. Scroll back and forth
in the program selection list with List DOWN|UP. Confirm the selected program line
by pressing Enter. The Enter key turns gray. If you would like to correct the program
memory location, select a new line number or highlight the new line directly on
the touch screen.
Re-confirm by pressing Enter. Press PAGE DOWN to scroll to the next page.
The diffusion test is used as an example.
If the optional Barcode Scanner is used for programming, please compare chapter
7.18 on page 49).
Optional Parameters for Detection of Disconnected Filters
In case of a disconnected filter (housing), an integrity test can be successfully
performed because the test-evaluating parameters (e.g. BP
min
, Diff
max
) are fulfilled.
To avoid such situations, it is possible to define four additional parameters.
A non-standard test (compare fig. 7-8) must be chosen to get access to these
parameters. As soon as at least one of these listed parameters are not fulfilled, a
warning message is displayed | printed. The following parameters can be defined:
Min. limits: Parameters can be activated | deactivated
The parameters can be activated / deactivated.
If deactivated the here below described limits will have no impact
on the test sequence.
Min. DIF/WIT/WFT: A minimal diffusion rate can be defined.
This parameter is valid for diffusion test, WIT and WFT.
If the measured value falls below this limit a warning message
will be printed at the end of the test together with the test
evaluation. It is then responsibility of the of the quality assurance
to decide if the test is conforming or not.
The test evaluation of the Sartocheck 4 plus remains “passed”.
Min. Flow BPT: A minimum flow rate can be defined for the Bubble Point Test.
This value represents the minimum flow that must be
measured at BP
max
.
If the measured value falls below this limit a warning message
will be printed at the end of the test together with the test
evaluation. It is then responsibility of the quality assurance to
decide if the test is conforming or not.
The test evaluation of the Sartocheck 4 plus remains “passed”.
Min. Volume: A minimal net volume can be defined. A net volume measure-
ment below this value will abort the test. An error message will
be printed.
Max. Volume: A maximal net volume can be defined. A net volume measure-
ment above this value will abort the test. An error message will
be printed.
Fig. 7-6 | Programming tests: Program selection
Fig. 7-6b | Optional parameters