EasyManua.ls Logo

Sartorius Sartocheck 4 plus - Test Methods; Programming Tests (F2 - Main Menu)

Sartorius Sartocheck 4 plus
187 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30 Operation
7.3.2 Test Methods With the Sartocheck
®
4 plus Filter tester, you can directly select a particular test
method. This greatly simplifies programming the data for a test, since only the
parameters that are relevant for the selected test are displayed.
You can select any of the following test methods:
Test methods
1 Diffusion test
2 Bubble point test (Standard and Customer specific)
3 Diffusion and bubble point test
4 Water intrusion test
5 Pressure drop test
6 Multipoint diffusion test (Standard and Customer specific)
7 Volume measurement
8 Plate holder test
9 Installed bag test
System Venting
At the end of each test, after a power failure or after switching off the unit during
a test, the system is automatically vented.
7.3.3 Programming Tests The Sartocheck
®
4 plus Filter tester cannot perform bag tests.
(F2 – Main Menu) Nevertheless it is possible to program bag tests due to the fact that additional Bag
tester MultiUnits may be connected to the Filer tester. The test parameters for the
individual tests are entered and selected on Sartocheck
®
4 plus Filter tester by using
the keypad or the touch screen. The Main Menu is the starting point for program-
ming. The following example uses a diffusion test to illustrate how to program
data for tests and run test sequences (F2 – Main Menu). The programming and test
sequences are identical for the other test types.
Before we start programming, we would like to explain a few important keys
and|or touch commands.
The Sartocheck
®
displays are set up with yellow buttons located on the right side
of the screen. These buttons always have the same functions:
Main Menu: The program jumps back to the Main Menu (see Fig. 7-5)
Page Up: the program scrolls to the previous page of the screen
Page Down: the program scrolls to the next page of the screen
List Up: Whenever a list is displayed on the screen, use LIST UP
to scroll up the list
List Down: Whenever a list is displayed on the screen, use LIST DOWN
to scroll down the list
Enter: Confirms the selected function, action or entry.
Following confirmation, the yellow key turns gray
Fig. 7-5 | Main Menu