8 Overview and Proper Use
2.3.5 Product-specific Data Manual input range upstream system volume 0
1
) –150000 ml
Measuring Ranges
– Test pressure: 100–8000 mbar g
– Pressure drop: 1–2000 mbar
– Max. measurable volume: 150 L
– Diffusion: 0.1–3000 ml/min
– Water intrusion: 0.1–999 ml/10 min
– Water flow: 0.01–99.9 ml/min
– Bubble point: 100–8000 mbar
Measuring Accuracy
– Pressure ± 0.1% | ± 9.5 mbar full scale
– Pressure drop ± 1 mbar
– Net volume measurement ± 4%
– Diffusion ± 5%
– Water intrusion ± 5%
– Bubble point ± 50 mbar
Operating Conditions
– Room temperature +15°C to +35°C
– Rel. humidity 10-80%
Touch screen (Sartocheck
®
4 plus Filter tester only)
– Size 10.4“
– Color 256 colors
Interface Port for Data Output
– Serial port TU RS232|RS485
– Serial port MU RS232 (Sartocheck
®
4 plus
Filter tester only)
– PLC port binary signals 12 pins
– Network Connection RJ45
Selectable languages:
English, German, French, Spanish, Italian
1)
When 0 is input, the upstream system volume is measured. Volume input and volume measurement are
not applicable for bag testing.