PLC Control 91
Name Function  Outlet Status
  D E F  G H L  M
Outlet D  Unit is ready to use (stand by) 1 0 0 0 0 0 0
Outlet E  Test running   1 1 0 0 0 1
1
 1
2
Outlet F  Error display  1 1 1 0 0 0 0
Outlet G  Test passed (OK)  1 1 0 1 0 0 0
Outlet H  Test failed (not OK)  1 1 0 0 1 0 0
Outlet L  Ext. valve 1  see note 1
Outlet M  Ext. valve 2  see note 2
Note 1: Outlet L not activated during a test until the end of stabilization time
Note 2: Outlet M not activated during a test until the end of measuring time
Plug Layout
Plug Function
Pin A  START
Pin B  STOP
Pin K  PROG
Pin C  GND
Pin D  POWER ON
Pin E  TEST RUNNING
Pin F  ERROR
Pin G  TEST OK
Pin H  TEST NOT OK
Pin J  +24 V DC
Pin L  EXT. VALVE 1
Pin M  EXT. VALVE 2
14.5 Outlets