Operation 57
7.19.1.3 Test unit address This function sets the address of the test unit so that the Sartocheck
®
4 plus Filter
tester can allocation recognize it. The address Test Unit #1 is reserved for the
Sartocheck
®
4 plus Filter tester.
Allowable addresses for the MultiUnits are Test Unit #2 through Test Unit #5.
MultiUnits come from the factory with no address allocation, and therefore
must be set by the user.
When connecting MultiUnits to the Sartocheck
®
4 plus Filter tester for the first time,
it is important that only one new MultiUnit be connected at a time. Two or more
units that have the same address allocation, or no address allocation (default setting
from the factory), will cause a communication conflict.
When transferring a MultiUnit to be used with a different Sartocheck
®
4 plus Filter
tester, it is advisable to first set the address allocation back to the default settings.
MultiUnit address allocation procedure:
t MultiUnit should be turned on, but the communication cable to the MultiUnit
should not be connected.
t From the Hardware configuration screen of the Sartocheck
®
4 plus Filter tester,
press F3 (Test Unit address allocation).
t Connect the MultiUnit communication cable when prompted by the
Sartocheck
®
4 plus Filter tester and press “Page down”.
t The new MultiUnit will be displayed in F7 as “New test unit” with the appropriate
serial number.
t Select F7 and press “Page down”.
t Available test unit address locations will be shown.
t Select the desired address location and press “Page down”.
Change a MultiUnits address allocation:
t From the Hardware configuration screen of the Sartocheck
®
4 plus Filter tester,
press F3 (Test Unit address allocation), then press “Page down”.
t Select the appropriate test unit and press “Page down”.
t Select the desired address location and press “Page down”.
To set the MultiUnit back to the factory default setting, select F7
[Reset to standard address] and press “Page down”.
Fig. 7-70 | Service Menu: Hardware configuration
Fig. 7-71 | Service Menu: Hardware configuration