78 Troubleshooting
Err no.: Type: Usertext:
9193 E Value above the max. volume.
9194 E Value below the min. volume.
9195 E The function was cancelled.
9196 W The internal temperature limit was exceeded.
9197 W The environmental temperature fluctuations during Stabilization time/Test time are out of the permissible tolerance of ±
<ARG1>. The actual temperature fluctuation is: Stabilization time: Min/Max ( <ARG2>) Test time: Min/Max ( <ARG3>)
9198 W The internal and external temperature limits were exceeded.
9199 W During monitoring using the pressure gauge function, the maximum allowed pressure was exceeded.
9201 E An internal valve has short-circuited at the supply voltage
9202 E An internal valve has short-circuited to ground
9203 E An internal valve has a broken conductor
9204 E An internal valve has short-circuited
9205 E An external valve has short-circuited at the supply voltage
9206 E An external valve has a broken conductor - Connect valve(s) - Electrically test valve(s)
9207 E The external filling valve has a broken conductor - Connect valve(s) - Electrically test valve(s)
9208 E The external drain valve has a broken conductor - Connect valve(s) - Electrically test valve(s)
9209 E The external vent valve has a broken conductor - Connect valve(s) - Electrically test valve(s)
9210 E The external block valve has a broken conductor - Connect valve(s) - Electrically test valve(s)
9211 E An external valve has short-circuited
9212 E An external proportional valve has short-circuited at the supply voltage
9213 E An external proportional valve has a broken conductor - Connect valve(s) - Electrically test valve(s)
9214 E The “small” external proportional valve has a broken conductor - Connect valve(s) - Electrically test valve(s)
9215 E The “large” external proportional valve has a broken conductor - Connect valve(s) - Electrically test valve(s)
9216 E An external proportional valve has short-circuited
9217 E An internal proportional valve has short-circuited at the supply voltage
9218 E An internal proportional valve has a broken conductor
9219 E The “small” internal proportional valve has a broken conductor
9220 E The “large” internal proportional valve has a broken conductor
9221 E An internal proportional valve has short-circuited
9222 E The external temperature sensor is not connected.
9301 E Internal error
9302 E internal error <ARG1>
9303 E Internal error
9304 E Valve has not been tested yet.
9305 E Valve does not close.
9306 E Valve does not open.
9307 E Valve 3 does not close. Function test should not be continued.
9308 E Valve 4 does not open. Function test should not be continued.
9309 ...
9310
E internal error <ARG1>
9315 E Communication error with A/D converter of the safety circuit
9316 E Communication error for reference adjustment of the safety circuit
9320 E The difference in pressure is too large. No meaningful adjustment possible.
9321 E The safety valve does not open although it has been electrically deactivated.
9322 E Adjustment could not be successfully completed within the time provided.
9323 ...
9325
E The safety valve opened because the pressure has been exceeded.
9326 E During test time, nominal pressure exceeded target pressure
9327 E The safety valve circuit did not respond although the pressure was too high.