16 Weigh Cell Operating Instructions
Installation
5.11 Maximum Load on the Pan Support
1
Fig. 6: Maximum load on the pan support
Pos. Name Description
1 Force holding point
Model
WZB25-NC
WZB254-N
WZB254-ND
WZB254-NC
WZB653-N
WZB1403-N
WZB1403-NC
WZB8202-N
WZB8202-NC
Unit Value Value Value Value Value
Maximum torques (Mx. My, Mz) Nm 0.1 2 2.5 4 8
Torque forces Nm 0.5 1 1 2 4
Maximum force opposite to direction of load (-F
z
) N 2 3 6 15 100
Maximum forces on the force holding point
(F
x
, F
y
, F
z
)
N 2 20 25 40 80
F
z
200
Either the maximum forces or torques may be present. If forces and torques occur at the same time, the sum of the
percentage loads must not exceed 100%.
Higher loads may destroy the weigh cell.
Example
Weigh cell with load holder protruding to the front.