20 Weigh Cell Operating Instructions
System Settings
7.6 Automatic Data Output
In “Automatic Printout” mode, data is output to the data
interface without an extra print command.
You can have synchronized data output automatically at
defined display update intervals, with or without weigh
cell stability. The time for an interval depends on the
setting under “Ambient conditions” (code 1. 1. 1. x) and
“Auto output cycle” (code 3. 1. 3. x).
If the automatic data output is activated in the menu, it
starts immediately after the balance is turned on. In the
setup, you can also specify whether the automatic data
output can be stopped and started with a “Print com-
mand” (key or via interface).
7.7 Adjustment
Procedure
t Perform adjustment functions:
— Via control commands with Sartorius
configuration software CAS Suite installed on a
computer (see Chapter “8.6 “Calibration,
Adjustment” Function via Interface RS232
(Examples)”, page 38)
— With the optional YRD01 display unit
7.8 Below-balance Weighing
There is a below-balance weighing unit for weighing
processes below the weigh cell (not for model
WZB25-NC).
Overload protection is available in the following models
as standard:
— WZB254-N | -ND
— WZB1403-N | -NC
— WZB653-N
— WZB8202-N |-NC
Procedure
t Remove the screw
plug (1).
1
7.5 Data Output Rates – Values per Second
Ambient Conditions
(Filter adaptation)
WZB...N|-NC WZB254-ND
XBPI SBI XBPI SBI
“Autoprint”
Very stable (1.1.1.1) 20 20 150 150
Stable (1.1.1.2) 10 10 150 150
Unstable (1.1.1.3) 5 5 150 150
Very unstable (1.1.1.4) 2.5 2.5 18.75 18.75