32 Weigh Cell Operating Instructions
Operation
8.1.4 Influence of Drafts
Depending on the size of the load holder and the sample, there may be an influence from drafts.
Use an appropriate draft shield to reduce the influence.
Protect the weigh cell from drafts.
8.2 Operation with a Computer
The weigh cells are equipped with two data interfaces (RS232 and USB-C), to which a computer or another peripheral
device can be connected.
You can use a computer to change, start, and monitor weigh cell functions.
Features
t See Chapter “13.3 Data Interfaces”, page 51.
Parameter Factory Settings
Transmission rate: 
Parity 
Stop bits 
Handshake

Communication mode Individual value without stability (3. 1. 1. 1)
Preparation
t See Chapter “13.3 Data Interfaces”, page 51 and Chapter “13.4 Cabling Diagram (Cable)”, page 51.
8.3 Output Format with 16 Characters (Compatibility with Current

Characters that are not visible in the display are output as spaces.
The type of character that can be output depends on the character's position
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ A A A A A A A * E E E CR LF
or - . . . . . .. . * * *
or * * * * * * * * *
* Space CR: Carriage return
A: Displayed characters LF: Line feed
E: Measurement unit characters .: Decimal point