Weigh Cell Operating Instructions 33
Operation
Special Outputs
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * * * * * * * * * * * * CR LF
or - H i g h
or L o w
or C a l . E x t .
* Space High: Overload
Cal.Ext.: Adjustment, external Low: Underweight
Error Message
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
E r r * # # # * * * * CR LF
A P P . E R R** * * * * CR LF
D I S . E R R** * * * * CR LF
P R T . E R R** * * * * CR LF
* Space ###: Error number
** For cause and solution, please refer to the “Error Codes” section
Example: Output of the weight value + 123.56 g
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ * * * 1 2 3 . 5 6 * g * * CR LF
+ * * 1 2 3 . 5 [ 6 ]** g * * CR LF
Position 1: “+” sign or space
Position 2: Space
Position 3 - 10: Weight value with decimal point; leading zeros are output as spaces
Position 11: Space
Position 12 - 14: Measurement unit characters or space
Position 15: Carriage return
Position 16: Line feed
** For cause and solution, please refer to the “Error Codes” section