Weigh Cell Operating Instructions 39
Operation
Performing Calibration and Adjustment Routines
The following settings can be configured:
Calibration and adjustment are always one process
(factory setting)
After calibration, the user has the option to quit the
process without correction or to make an adjustment.
If no deviations are found during calibration, the
calibration/adjustment process can be exited after the
calibration is completed.
Two keys are now active:
Starting the adjustment: ESC f1_ command
Exiting the process: ESC s3_ command
Procedure
t To tare the balance: Select [ESC T] interface.
y The display shows 0.0000 g.
t To start the adjustment process: Select [ESC Z]
interface.
y The display shows Cal.INT.
y The internal weight is applied automatically.
y The display shows CALRUN.
y Adjustment is executed.
y The display shows CAL.END.
y The internal weight is removed from the balance
y The display shows 0.0000 g.
Model WZB25-NC
Requirement
The application SETUP - Balance - CAL./ADJ. - Ext.
adjustment with user-defined weight is set up.
The application SETUP - Balance - Adjustment
process - Adjust immediately is set up.
The weigh cell is unloaded.
Procedure
t To start the adjustment process:
t If the compatibility with older weigh cells is active:
Select the ESC s1_ interface.
t If the compatibility with older weigh cells is not
active: Select ESC f1_ interface.
y The prompt to apply the adjustment weight appears.
t Apply pressure to the pressure inlet of the internal

t Wait until the weight value is adopted.
t Switch off the pressure.
8.6.2 External Adjustment
Requirement
The application SETUP - Bal.Scal - CAL./ADJ. -
Cal.EXT. is set up.
A weight value is preset at the factory (see
Chapter “13 Technical Data”, page 50).
Procedure
t To tare the balance: Select [ESC T] interface.
y The display shows 0.0000 g.
t To start the adjustment process: Select [ESC W]
interface.
y The display shows Cal.EXT.
y Once you store the zero point, there is a flashing
display prompt for the adjustment weight to be
applied:
- 50.0000 g.
t Apply the displayed adjustment weight (in this

y If the “-” sign is displayed: The weight is too small.
y If the “+” sign is displayed: The weight is too big.
y If the adjustment weight is within the specified limits:
The flashing stops.
y If the adjustment has been carried out:
y The display shows Cal.END.
y The adjustment weight appears: +50.0000 g.
t Remove the adjustment weight.
y The display shows 0.0000 g.