Weigh Cell Operating Instructions 45
Operation
8.8.1 Switching YRD01 On and Off
Procedure
t
CAUTION Pointed or sharp-edged objects may
damage the operating display!
t Only touch the operating display with your fingertips.
t To switch the device on: Press the [On/Off] button.
t To switch the device off: Press and hold the [On/Off]
button.
8.8.2 Calibrating and Adjusting Device with
isoCAL Function
Triggers for the information on the isoCAL function are:
The ambient temperature has changed since the last
adjustment process.
The interval time has been exceeded.
Requirements
The information on the isoCAL function is set in the
menu.
The device is not located in the menu settings.
Procedure
t If the [isoCAL] button is flashing in the operating
display: Press the [isoCAL] button.
y The isoCAL function starts.
8.8.3 Internally Calibrating and Adjusting the
Device
Requirements
The weighing pan is unloaded.
The operating display shows a stable weight value.
The internal calibration and adjustment function is
set up.
Procedure
t Press the [Zero] button.
t Press the [Adjust] button.
y The current calibration | adjustment function is
displayed.
t If the calibration with subsequent automatic
adjustment function is selected:
y The “CAL.RUN.” display appears in the operating
display during the process.
y The “CAL.END” display indicates the end of the
adjustment process.
t If the calibration without subsequent automatic
adjustment function is selected:
y The calibration process is started.
y The “CAL.RUN.” display appears in the operating
display during the process.
y The calibration error is displayed.
t Press the [Confirm] button
y The adjustment process is performed.
y The “CAL.END” display indicates the end of the
adjustment process.
8.8.4 Externally Calibrating and Adjusting
the Device
Material Calibration and adjustment weight
Requirements
The weighing pan is unloaded.
The operating display shows a stable weight value.
The external calibration and adjustment function is
set.
The preset weight value for the adjustment weight or
the user-defined weight value for the adjustment
weight is selected.
Procedure
t Press the [Zero] button.
t Press the [Adjust] button.
t Place the indicated calibration and adjustment
weight on the balance.
y The weight value is displayed.
t If the calibration with subsequent automatic
adjustment function is selected and the adjustment
weight placed on the device is within the specified
limits:
y The calibration | adjustment process starts.
y If the “+” sign is displayed: The applied weight is
too big.
y If the “-” sign is displayed: The applied weight is
too small.