EasyManua.ls Logo

Schiller ARGUS LCM PLUS - Procedure in Case of an Alarm; Display of Alarms; Switching off an Alarm

Default Icon
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3Operation
3.7 Procedure in case of an alarm
Page 22
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
3.7 Procedure in case of an alarm
3.7.1 Display of alarms
During initial switching on
The alarms are suppressed for a defined time (programmable in the menu About+/
Alarm. Alarm suppr. time standard is 3 minutes)
The message Alarm suppressed 3:00 appears.
During monitoring
There are two alarms:
Technical alarm, displayed in the alarm status field on the top right.
In the case of a technical alarm, an audible alarm sounds and the measurement
field for the respective value flashes.
Physiological alarms, displayed in the alarm status field on the top left.
In the case of a physiological alarm, an acoustic alarm sounds and the measure-
ment field for the respective value flashes.
3.7.2 Switching off an alarm
Each alarm off function is password protected. Read carefully the following warning
and information:
Entering alarm OFF password
V When the acoustic alarm is switched off, the patient acoustic physiological
alarms are silenced and suppressed indefinitely. Use this function only if discon-
necting a sensor from the patient for a long period of time.
When the alarm off button is pressed, the audible alarm for the respective param-
eter is suppressed. This is indicated by the LED above the button and a message
on the monitor, e.g. Alarm ECG OFF. If an alarm occurs as long as the alarm off
button is pressed, a visual alarm is displayed in the respective measurement field.
With the LCM plus version E the access to the NIBP and IBP alarm off function will
be executed by the NIBP alarm button. A menu with blood pressure alarm off (com-
mon for NIBP and IBP) and IBP calibration function appears.
BP Alarm On/Off
Connect zero mmHg
Start IBP calibration
End / Escape
1. Press the desired alarm off button. The password dialog appears.
2. Press following button to enter the password:
, , ,
1. The password protection can be disabled in the special menu see chapter 7.1.2
page 44
Enter Password
Press [Enter] to cancel
I,II, II
I

Table of Contents

Related product manuals