Page 45
Appendix 7
Operating Instruction Factory settings 7.2
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
7.2 Factory settings
These values can be loaded from the menu About+/Software. See chapter 7.1.2.
Parameter Abbrev. Unit Type Limit Step Alarm
Prio.
Adult Child Neonat
Heart rate HR min High 80...220 5 Low 140 180 200
min Low 15...100 5 50 70 90
Asystole s 2...20 1 High 2 2 2
Respiration RR min High 25...120 5 Low 35 35 60
min Low 1...20 5 6 12 20
Apnea s 15...45 1 Low 20 15 10
NIBP SYS mmHg High 80...300 5 Low 180 180 80
mmHg Low 10...150 5 100 100 50
DIA mmHg High 30...120 5 Low 95 95 45
mmHg Low 0...80 5 40 40 30
MEAN mmHg High 0...200 5 Low 140 140 45
mmHg Low 0...100 5 70 70 30
Pulse min High 80...220 5 Low 140 180 200
min Low 15...100 5 50 70 90
Pulsoximetry
SpO
2
% High 80...101 1 High 101 101 98
% Low 70...99 1 90 89 88
PP min High 80...220 5 Low 140 180 200
min Low 15...100 5 50 70 90
Capnography
eC0
2
Vol% High 4.5...10 0.1 Low 6.0 6.0 6.0
Vol% Low 0...6.0 0.1 4.5 4.5 4.5
CO
2
i
Vol% High 0.1...1.5 0.1 Low 0.2 0.2 0.2
Vol% Low 0...1 0.1 0 0 0
eCO
2
kPa High 4.5...10 0.1 Low 6.0 6.0 6.0
kPa Low 0...6.0 0.1 4.5 4.5 4.5
CO
2
i
kPa High 0.1...1.5 0.1 Low 0.2 0.2 0.2
kPa Low 0...1 0.1 0 0 0
eCO
2
mmHg High 35....76.0 0.5 Low 45.0 45.0 45.0
mmHg Low 0...45.0 0.5 35.0 35.0 35.0
CO
2
i
mmHg High 1...10.0 0.5 Low 1.5 1.5 1.5
mmHg Low 0...8.0 0.5 0 0 0
IBP Ps mmHg High 0...300 1 Low 180 180 180
mmHg Low -20...150 1 95 95 95
Pm mmHg High 0...200 1 Low 100 100 100
mmHg Low -20...150 1 50 40 40
Pd mmHg High 0...120 1 Low 100 100 100
mmHg Low -20...80 1 40 40 40
T
1
T
1
°C or °F No alarm No settings - - - - -