6 Technical Data
6.2 Technical data - measured values
Page 42
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
6.2.6 SpO
2
- pulsoximetry
6.2.7 etCO
2
- capnography
Amplifier
Masimo™ MS-3, MS-7, NELL-1
Operation
Normal and sensitive
Sampling rate
62.5 Hz
Accuracy
SpO
2
Adults 70 to 100% ± 2 digits
Neonates 70 to 100% ± 3 digits
PP
30 ... 199/min ± 4 digits
Calibrated range
70 ... 100% (calibraton is fixed, no calibration required)
Measurement range
SpO
2
1 ... 100%
PP 25 ... 240/min
Displayed range
1 ... 100%
PP calculation
8 s
Measuring method
Side stream
Displayed range
0 ... 99 mmHg
Suction rate
90/120/200 ml/min
Accuracy
± 3 mmHg at 0 ... 40 mmHg
± 8 mmHg at 41 ... 76 mmHg
± 10 mmHg at 77 ... 99 mmHg
Respiration rate
0 ... 99 inspirations/min
Environmental pressure com-
pensation
Automatic when the unit is switched on and when the watertrap is inserted.