Technical Data 6
Operating Instruction Technical data - measured values 6.2
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
6.2.3 Temperature
6.2.4 NIBP - non-invasive blood pressure
6.2.5 IBP - invasive blood pressure
Sensor
YSI 401, rectal, skin or ear
Measurement interval
1x per second
Measurement range
15 °C to 45 °C
Resolution
0.1 °C
Minimum measurement
duration
2 min until a measured value is achieved
Measurement
Automatic or manual
Measuring method
Oscillometric
Measurement range
15 to 300 mmHg
Accuracy
Max. mean error
Max. standard deviation
± 5 mmHg
± 8 mmHg
Deflation rate
3 to 9 mmHg
Measurement range
-20 ... 300 mmHg
Sampling rate
500 Hz
Accuracy
1 mmHg at 0...100 mmHg or
±1% at 100...300 mmHg
other sensors may cause lower accuracy
Calibration
Manual
Pulse calculation
8 beats