EasyManua.ls Logo

Schiller ARGUS LCM PLUS - SpO2 Monitoring; IBP Monitoring; CO2 Monitoring

Default Icon
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 5
Safety notes 1
Operating Instruction Responsibility of the user 1.1
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
1 Safety notes
1.1 Responsibility of the user
1.2 Intended use
1.3 Organisational measures
V This device must only be used by qualified doctors or trained medical personnel
under their direct supervision.
V The numerical and graphical results and any interpretation given must be exam-
ined with respect to the overall clinical condition of the patient and the general re-
corded data quality.
V The indications given by this equipment are not a substitute for regular checking
of vital functions.
V Specify the competencies of the personnel for operation and repair.
V Ensure that the personnel have read and understood these operating instructions
and in particular this chapter “safety notes".
V Have damaged or missing components replaced immediately.
V The operator is responsible for compliance with all applicable accident prevention
and safety regulations.
V The ARGUS LCM/PLUS is a patient monitoring device used for the measuring of
the parameters of a patient, including ECG, SpO
2
, CO
2,
non invasive blood pres-
sure, temperature and respiration.
V The system is suitable for internal hospital transport.
V There is no danger for patients with pacemaker.
V The device is only intended for single patient use.
V Only operate the device in accordance with the specified technical data.
V The system is not designed for sterile use nor is it designed for outdoor use.
V Do not use this unit in areas where there is any danger of explosion or in the pres-
ence of flammable gases such as anaesthetic agents.
V This unit is CF classified. It is defibrillation protected only when the
SCHILLER original patient cable is used. However, as a safety precaution when
possible, remove electrodes before defibrillation.
V This product is not designed for direct cardiac application.
V Before using the unit, ensure that an introduction regarding the unit functions and
the safety precautions has been provided by a medical product representative.
V Keep these operating instructions in an accessible place for reference when re-
quired. Make sure that they are always complete and legible.
V Observe the operating instructions and maintenance instructions.
V These operating instructions do not override any statutory or local regulations, or
procedures for the prevention of accidents and environmental protection.

Table of Contents

Related product manuals