EasyManua.ls Logo

Schiller ARGUS LCM PLUS - Zero and Two Point CO Calibration; Starting the Zero Point Calibration; Starting the Two Point Calibration

Default Icon
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 35
Maintenance 5
Operating Instruction Zero and two point CO
2
calibration 5.5
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
5.5 Zero and two point CO
2
calibration
5.5.1 Starting the zero point calibration
5.5.2 Starting the two point calibration
•CO
2
scrubber not more than one year old.
Watertrap
Calibration gas 10% CO
2
Bal. N
2
Or calibration kit including above items and accessory Art.no. 2.100741
Calibrate the sensor's zero point voltage at dry, clean air and at room temperature.
Zero point calibration must be carried out every two weeks. A zero point calibration
is successful if the CO
2
value is in the range of 0.0% to 0.3%.
Two point calibration must be carried out every six months. A two point calibration
automatically includes zero-point calibration. A two point calibration is successful
if the CO
2
value is 10% after one minute.
The calibration date is saved by the LCM and displayed in the CO
2
settings panel.
When the calibration interval has elapsed, a message is displayed.
To reduce the calibration time, you can change the respiration source to CO
2
in the
menu About+/Param. The pump starts as soon as the menu is closed.
1. Connect the watertrap (1) to the CO
2
socket.
2. In the CO
2
screen, select Start zero point calibration and confirm with enter.
The pump is started.
3. Continue with enter. You are prompted to connect the CO
2
scrubber.
4. Connect the CO
2
scrubber (2) to the watertrap (1).
5. Press enter to continue. The pump must run five min. before the calibration is car-
ried out automatically.
6. When the calibration is finished, a message is displayed. Finish the calibration by
pressing enter.
1. Connect the watertrap (1) to the CO
2
socket. (See zero point calibration.)
2. In the CO
2
screen, select Start two point calibration and confirm with enter. The
pump is started.
3. Continue with enter. You are prompted to connect the CO
2
scrubber.
4. Connect the CO
2
scrubber (2) to the watertrap (1).
5. Press enter to continue. The pump must run five min. before the calibration is car-
ried out automatically. When the zero point calibration is finished, you are prompt-
ed to connect the calibration gas.
6. Remove the CO
2
scrubber and connect the calibration gas (3) to the watertrap (4)
using a T-piece (5).
7. Open the valve of the gas bottle. An eCO
2
value of about 10% is indicated, and
the calibration is carried out.
8. When the calibration is finished, a message is displayed. Finish the calibration by
pressing enter.
1
2
3
5
4

Table of Contents

Related product manuals