3Operation
3.7 Procedure in case of an alarm
Page 24
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
3.7.4 Preventive alarm suppression
The preventive alarm suppression is used to deactivate in advance all alarms that
may be caused by disconnecting patient cables, lose electrodes and relocation of
the patient.
3.7.5 Overview of physiological alarms
« Press the alarm button and confirm with the enter button before an alarm is dis-
played.
Message Alarm suppressed 3:00 is displayed. The time can be programmed in the
menu About+/Alarm/Alarm suppr. time.
The alarms are reactivated after the defined period of time has elapsed.
Removing the patient cables
When a cable is removed, the message Cable off or no Sensor is displayed.
« Press the alarm suppressing button.
The alarm is deleted and the measurement and wave field is no longer displayed.
Alarm abbreviation Description
Asys limit Asystole time limit exceeded
SpO
2
Low/High
Oxygen saturation of the blood
PP Low/High
Peripheral pulse of SpO
2
RRi Low/High Respiration rate impedance (from ECG electrode)
Apnea limit Apnea time limit exceeded
CO
2i
low/high Inspiratory CO
2
RRc low/high Capnographic respiration rate
eCO
2
low/high End-tidal expiratory CO
2
SYS low/high Systolic pressure
DIA low/high Diastolic pressure
MAD low/high Mean atrial pressure
HR low/high Heart reat
Ps low/high Invasive systolic blood pressure
Pm low/high Invasive mean blood pressure
Pd low/high Invasive diastolic blood pressure
Temp low/high
When the temperature is outside the measuring range,
this is indicated by “<<“(15 below °C) or “>>” (above 45
°C). This limit is fixed.